System
Automatic high-pressure steam sterilization
Operating temp. range
45~135℃(45~80℃[pre heating temperature]/45~60℃[heat retention phase ]/65~100℃[disolution phase ]/105~135℃[sterilizing phase])
Max. working pressure
0.255MPa
Ambient temperature
5~35℃
Lid structure
Manual upward opening with an interlock for safety
Exhaust valve
One exhaust valve and one slow release valve
Connection ports for optional accessories
Total 3 ports. Female thread for sample temp. sensor (1/4" ), Female thread for chamber temp. sensor (1/4" ),
Cooling fan
Axial fan motor
Temp. control method
PID control by microprocessor
Temp. display / setting
Digital display / digital setting by ▲/▼ keys
Timer
0 or 1 min. to 99 hrs. 59 min. (resolution: 1 min.)
Operation functions
Instrument sterilization course, Liquid sterilization course, Sterilization and keep warm course, Melting and keep warm course,
Other functions
Key locking, Presetting, Saving, preheating, Forced cooling, Sample temperature sensor (option), Pattern locking,
Safety functions
Sensor failure detection, SSR short-circuit, Broken heater wire, Prevention of idle heating (Liquid expansion type),
Accessories
2 pcs. stainless steel mesh basket, Vapor cup×1, Drain bottle×1, Drain board×1, Filter×1
Internal dimensions (diameter×height mm)
370×442
External dimensions (W×D×H mm)
520×660×846
Power supply
AC200~240V 12.5~10.5A